Reports of 58548
AdV-TCS (CO2 laser projector)
lumaca - 12:14 Thursday 02 May 2024 (64156) Print this report
ONCALL - WI CO2 DAS OUT DMS flag

Around 08.50 UTC Francesco called me from CR, because the DMS flag of WI_CO2_Laser became red for mean_TCS_WI_CO2_PWROUT value falling down.

First of all, I recovered the WI DAS outer ring power injected inside the interferometer, rotating the corresponding waveplate (action ended around 08.57 UTC)

Then, I try to recover WI CO2 main laser state by acting on its chiller setpoint:

  • at 09.00 UTC WI e BCK chiller setpoint changed from 19.03°C to 19.02°C → wrong direction

  • 09.05 UTC WI e BCK chiller setpoint changed from 19.02°C to 19.04°C

Once the situation was re-established, I also recovered the  waveplate (action ended at 09.20 UTC).

In Fig. 1 the main signals are reported.

Images attached to this report
Virgo Runs (O4b)
menzione - 6:52 Thursday 02 May 2024 (64154) Print this report
Operator Report - Night shift

I found ITF in SCIENCE MODE.
It remained locked for the whole shift except for one unlock, at 21:22 UTC, done by mistake.

Guard Tour (UTC):
21:06 - 21:36
23:05 - 23:35
01:00 - 01:30
03:02 - 03:32 

Images attached to this report
Virgo Runs (O4b)
Sposito - 22:57 Wednesday 01 May 2024 (64153) Print this report
Operator Report - Afternoon shift

I found ITF in SCIENCE. 15:32 UTC ITF in CALIBRATION to perform the scheduled activity.

  • 15:33-16:33 UTC CALIBRATED_DF_SENSITIVITY.
  • 16:34-17:09 UTC CALIBRATED_DF_PCAL 
  • 17:10-17:22 UTC CALIBRATED_DF_DAILY

17:30 UTC CALIBRATION ended, back to SCIENCE

 

Images attached to this report
Virgo Runs (O4b)
berni - 15:02 Wednesday 01 May 2024 (64152) Print this report
Operator Report - Morning shift

ITF found in relocking; it relocked at 5:35 UTC and it remained in Science mode for the rest of the shift.

 

Guard tours (time in UTC)

From 6:00 to 6:30; from 8:00 to 8:30; from 10:00 to 10:30; from 12:00 to 12:30.

 

Images attached to this report
Virgo Runs (O4b)
ruggi - 11:47 Wednesday 01 May 2024 (64150) Print this report
Comment to Operator Report - Night shift (64147)

The earthwquake produced an increase of the seismic component at 1 Hz, which is a problem since the MAR/MIR actuation splitting filters have been changed: the unlock occurred when NE MIR CORR saturated (fig 1). Usually the reason of unlock during an earthquake is the saturation of MAR correction, due to the low frequency component of the surface seismic waves. A magnitude 4 in Croatia is not strong enough to produce a large slow oscillation of the suspension able to push the MAR corrections close to 10 V (fig 2). This kind of unlock would be likely avoided finding a way to restore the old MAR/MIR allocation strategy.

Images attached to this comment
Virgo Runs (O4b)
berni - 9:21 Wednesday 01 May 2024 (64149) Print this report
Comment to Operator Report - Night shift (64147)
Virgo Runs (O4b)
mwas - 7:55 Wednesday 01 May 2024 (64148) Print this report
Comment to Operator Report - Night shift (64147)

The unlock at 3:41 UTC could have been due to an earthquake. There was a seismic disturbance in both NEB and WEB at about that time.

Images attached to this comment
Virgo Runs (O4b)
amagazzu - 6:52 Wednesday 01 May 2024 (64147) Print this report
Operator Report - Night shift

ITF found in LOW_NOISE_3_SQZ and in Science Mode. The ITF unlocked at 3:41 UTC, and failed the relock twice at LOCKING_CARM_NULL_3F (3:57 UTC) and ACQUIRE_LOW_NOISE_1 (4:34 UTC). 
Relock in progress.

No DMS alerts received during the shift.

Guard Tour (UTC):
21:06 - 21:33
23:00 - 23:30
21:34 - 22:03
03:45 - 04:17

Images attached to this report
Comments to this report:
mwas - 7:55 Wednesday 01 May 2024 (64148) Print this report

The unlock at 3:41 UTC could have been due to an earthquake. There was a seismic disturbance in both NEB and WEB at about that time.

Images attached to this comment
berni - 9:21 Wednesday 01 May 2024 (64149) Print this report
ruggi - 11:47 Wednesday 01 May 2024 (64150) Print this report

The earthwquake produced an increase of the seismic component at 1 Hz, which is a problem since the MAR/MIR actuation splitting filters have been changed: the unlock occurred when NE MIR CORR saturated (fig 1). Usually the reason of unlock during an earthquake is the saturation of MAR correction, due to the low frequency component of the surface seismic waves. A magnitude 4 in Croatia is not strong enough to produce a large slow oscillation of the suspension able to push the MAR corrections close to 10 V (fig 2). This kind of unlock would be likely avoided finding a way to restore the old MAR/MIR allocation strategy.

Images attached to this comment
AdV-SGD (FDS commissioning)
sorrentino, demarco, lartaux - 0:22 Wednesday 01 May 2024 (64146) Print this report
SQZ injections vs SR alignment

We started at 14:00 UTC and to not interfere with the ITF and risk to unlock, we manually injected the SQZ. 

  • System initial status: ITF LOW NOISE 3; SQZ_MAIN @SQZ_Locked_NO_FC

  • 14:28:26 AA loop engaged

 

We performed a first CC phase scan (Fig. 1-2): 

  • 14:39 CC scan started DCP = 195 Hz

  • 15:02:21 CC scan ended (fast shutter closed), ITF in shot

  • 15:26:15 fast shutter opened

  • 15:36:22 ITF in SQZ phi0 = 0.65 rad

  • 15:39:19 SQB1 TX angle set to 300 (initially at 310)

  • 16:42:58 SQB1 X position set to -3200 (original position)

  • 16:47:58 ITF in SQZ

 

After the first CC scan we realized the magnitude was similar to the last shift (entry #64094) even though there should be a higher level of injected SQZ. We decided to try to improve the magnitude by acting on SQB1_TX and SQB1_X DOFs. We worked on this from 15:39:19 to 16:42:58 UTC. Finally we left SQB1_X as it was and we changed the set point of SQB1_TX from 310 to 330 urad. The fluctuations on the magnitude are reduced (Fig. 3).

We then performed another phase scan (Fig. 4-5)

  • 17:05 CC scan started DCP = 195 Hz

  • 17:28 CC scan ended (fast shutter closed), ITF in shot

 

After the second CC scan we realized the SQZ angle is actually around 1.05 rad, so we decided to repeat the SQZ mode acquisition.

  • 17:39:53 fast shutter opened, ITF in SQZ phi0 = 1.05 rad

  • 17:49:08 CC loop engaged, gain=75000 (glitch between 17:54:50 and 17:56:14)

  • 17:58:54 ITF in ASQZ phi0 = 2.75 rad

 

We were able to get another CC phase scan with DCP setpoint at 225 Hz (Fig. 6-7). Fig. 8 reports also the shot SQZ and ASQZ acquisitions made after the scan.

  • 18:11 DCP value set to 225 Hz

  • 18:44:11 ITF in shot

  • 18:46: CC scan started DCP = 225 Hz

  • 19:09 CC scan ended

  • 19:15:07 ITF in ASQZ phi0 = 2.7 rad

  • 19:22:48 ITF in SQZ phi0 = 0.95 rad

  • 19:28 DCP value set back to 195 Hz

Images attached to this report
Virgo Runs (O4b)
Montanari - 23:04 Tuesday 30 April 2024 (64143) Print this report
Operator Report - Afternoon shift

List of activities and actions performed during the shift:

  • At the beginning of the shift (13:09UTC), ITF unlocked due to the increase of the power on B1 photodiode and consequently the close of the fast shutter (first plot attached)
  • Relock at first attempt, and coming back to SCIENCE at 13:54UTC
  • From 14:04UTC to 14:UTC, ITF in CALIBRATION Mode to check hrec as planned (CALIBRATED_DF_DAILY)
  • After that, started planned SQZ activities, so I put ITF in COMMISSIONING Mode at 14:19UTC
  • 14:48UTC VPM restarted
  • 19:41UTC stopped COMMISSIONING and I put back ITF in SCIENCE
  • 20:07UTC I disengaged the squeezing and exit from SCIENCE to allow H. Valbruch to solve system issues of this morning
  • 20:20UTC SCIENCE Mode

 

Guard Tour (UTC)
17:08 - 17:37
19:06 - 19:32

Sub-system reports

Vacuum
16:15UTC lost of signals from 600W Tube Station
I notified A. Pasqualetti who agreed to carried out checks probably not before next maintenance.
In according with him I shelved related flags until Thursday

Images attached to this report
AdV-ISC (LSC Noise Budget)
mwas - 20:12 Tuesday 30 April 2024 (64145) Print this report
Comment to LSC noise injections results (64132)

Figure 1. The CARM noise injection caused SR to move by ~0.5urad, because it made the B1 photodiodes saturate. It needs to be understood how the CARM noise injection shape needs to be changed so it doesn't cause B1 saturations which will spoil the measurement.

Figure 2. Normal daily hrec calibration measurements (between 16:00 UTC and 16:10 UTC)  are no longer kicking the SR alignment.

Images attached to this comment
Infrastructures (Air Conditioning)
soldani - 16:37 Tuesday 30 April 2024 (64144) Print this report
HVAC tuesday maintenance_30.04.2024
During Tuesday maintenance, this morning:
- carried out weekly checks for boiler, AHU and chillers in TB, CEB and MC
- refill of the LPG tank in Central area from 8.30 to 8.45 (UTC)
- carried out periodic check for AHU fan-motor belts state of tension and consunption (with external firm) with a short shut down of the involved machines:

AHU Clean rooms CEB: off at 6.40 - on 7.25 (UTC) ---> status OK
AHU INJ: off at 7.30 - on at 7.45 (UTC) ---> status OK
AHU DET: off at 8.53 - on at 9.08 (UTC) ---> status OK
Virgo Runs (O4b)
Sposito - 15:01 Tuesday 30 April 2024 (64135) Print this report
Operator Report - Morning shift

ITF is found in SCIENCE. 

At 6:01 UTC I set the ITF in Maintenance Mode, here is the list of the activities communicated to the Control Room:

Periodic Tasks carried out by the Operator:

  • From 6:19 UTC to 6:36 UTC, Check of IMC Working Point and Position;
  • 07:04-07:09 UTC OMC locked in SINGLE_BOUNCE_NI
  • From 7:11 UTC to 7:32 UTC - OMC Scan in Single Bounce configuration;
  • At 8:00 UTC - TCS Thermal Camera Reference;
  • At around 06:23 - 06:35 UTC - TCS Power Checks (see table below for results):
  CH (W) Inner DAS (W)  Outer DAS (W)
W Pickoff 0.255 0 0.22
N Pickoff 0.68 0.065 0.602

11:25 UTC Maintenance ended, I re-aligned the CITF and relocked the ITF at the first attempt, back in SCIENCE at 11:35 UTC.

11:58 UTC Vpm restarted

SQZ not working properly, F. Sorrentino contacted H. Valbruch for it.

Sub-system reports

ISC
I changed the set-points of NI and WI etalon suggested by M. Mantovani.
12:13 UTC WI new set point 19.825 and NI new set point 20.1

Images attached to this report
AdV-TCS (CO2 laser projector)
lumaca - 14:54 Tuesday 30 April 2024 (64141) Print this report
CO2 power check

[entry unexpectedly stuck in drafts]

During the night the WI CO2 laser changed the state, but ITF remained locked: Magazzu sent an email and report about the DMS flag in the Virgo logbook (64134).

During this morning maintenance activities were foreseen, so I waited at least for the taking of thermal camera images and chiller refilling before making actions for WI CO2 laser power recovery. After the chiller refilling and filter change, the WI CO2 main laser state was mainly recovered (see Fig. 1).

 

A further activity was foreseen during maintenance in TCS room (repair WI etalon heating belt, 64114). I waited for the end of this activity to further check CO2 powers (Gianmatteo already checked them 64135) and eventually act on waveplates.

At 09.00 UTC, with ITF in down, I checked all CO2 powers restoring the nominal ones. This activity has been completed at 09.18 UTC.

Images attached to this report
Environmental Monitoring (Environmental Monitoring)
fiori, graziaplena - 13:41 Tuesday 30 April 2024 (64142) Print this report
temporary magnetic probe
Vacuum (Detection)
Vacuum - 11:40 Tuesday 30 April 2024 (64140) Print this report
DT turbo pump cooling fan replacement

This morning during the maintenance break, we replaced a fan on the DT turbomolecular pump after noticing some unusual noise, possibly due to a worn bearing. We installed a standard domestic fan, mounted on rubber pads as usual (picture attached)

Images attached to this report
Detector Characterisation (Glitches)
dattilo, cavalieri - 11:35 Tuesday 30 April 2024 (64139) Print this report
Comment to New numerous glitches starting last night (64114)
As a follow up of the recent investigations on the WI etalon problem, we made a check of the WI heating belt power supply.
At around 08:30 UTC we went in TCS room where the power supply is located, and also looking at its analog display, we saw that it was delivering intermittently, as indicated by the LSC_WI_HB_moni signal.
We replaced the power supply with a spare one of the same kind (Kert 420). During the replacement (from 08:34 to 08:40 UTC) the operator temporarily opened the WI etalon loop, as recommended by Maddalena.
Now the loop seems working regularly (attached fig).
Images attached to this comment
AdV-SVS (Pre-commissioning)
sorrentino - 11:28 Tuesday 30 April 2024 (64137) Print this report
Improving MZI and OPA alignment in SQZ board

This morning I slightly tweaked the alignment of the optics showed in entry 56246 inside the AEI sqzueezer board, in order to improve the MZI contrast and the pump beam alignment on the OPA. 

I worked on MZI between 7:43 and 7:47 UTC, and on OPA between 7:52 and 8:19 UTC.

The MZI at the beginning was already quite high, and I could only improve it marginally  - from 68% to 73%, see first attached plot.

After tweaking the pump beam alignment, the OPA peak trasmission improved by about 5% (see second attached plot) and the main HOM in OPA scan was reduced by a comparable amount (see third attached plot).

 

 

Images attached to this report
AdV-SBE (Minitowers)
bulten, was, bertolini, bersanetti, gouaty, arnaud - 10:47 Tuesday 30 April 2024 (64138) Print this report
DMS flags added to horizontal positions SBE benches
When the horizontal microsprings are adjusted, or maybe when other horizontal actions on the minitower benches are performed, the DMS system does not give a warning if the position is still far off the setpoint. This happened April 16, see logbook entry 64002 from Nicolas Arnaud. After discussions, we decided to add flags to the monitoring system; we had flags for the vertical position (-30that for the relative speeds for SNEB and SWEB with respect to the superattenuators, but this requires further study to see if there are cases that the relative speed between benches is high, but none of the flags (RMS actuator voltage, bench error signal) is already triggering.
IT infrastructure (Computing)
kraja, cortese - 9:56 Tuesday 30 April 2024 (64136) Print this report
Firmware upgrade on underlying virtual infrastructure
During this maintenance window we will perform some necessary upgrades on the underlaying virtual infrastructure where the fileserver are hosted.
This operation is totally transparent for the hosted systems on the infrastructure, and should not impact the fileservers at all.
Virgo Runs (O4b)
amagazzu - 7:01 Tuesday 30 April 2024 (64134) Print this report
Operator Report - Night shift

ITF found in LOW_NOISE_3_SQZ and in Science Mode. It kept the lock for the whole shift.

Guard Tour (UTC):
21:01 - 21:29
23:02 - 23:29
01:01 - 01:29
03:01 - 03:29

Sub-system reports

TCS
At around 4:25 UTC the value of TCS_WI_CO2_PWROUT suddenly decreased (see attached plot). Expert informed via mail of the issue.

Images attached to this report
Virgo Runs (O4b)
menzione - 22:57 Monday 29 April 2024 (64133) Print this report
Operator Report - Afternoon shift

I found ITF in SCIENCE MODE; It remained locked for the whole shift. 

No DMS Events to report

Guard Tour (UTC): 
17:08 - 17:35 
19:01 - 19:29

Images attached to this report
AdV-ISC (LSC Noise Budget)
boldrini, pinto - 15:47 Monday 29 April 2024 (64132) Print this report
LSC noise injections results

The automated noise injections on Friday 26th worked as intended: there were two minutes of noise injection on each longitudinal DoFs (Figs.1-5).

The coherence with Hrec is unsatisfying though (Figs.6-10). A finer tuning of the amplitudes of the noise injections might be necessary.

Images attached to this report
Comments to this report:
mwas - 20:12 Tuesday 30 April 2024 (64145) Print this report

Figure 1. The CARM noise injection caused SR to move by ~0.5urad, because it made the B1 photodiodes saturate. It needs to be understood how the CARM noise injection shape needs to be changed so it doesn't cause B1 saturations which will spoil the measurement.

Figure 2. Normal daily hrec calibration measurements (between 16:00 UTC and 16:10 UTC)  are no longer kicking the SR alignment.

Images attached to this comment
Virgo Runs (O4b)
Sposito - 14:57 Monday 29 April 2024 (64131) Print this report
Operator Report - Morning shift

I found ITF in SCIENCE; It remained locked for the whole shift.

Images attached to this report
Detector Characterisation (Spectral lines)
direnzo - 11:57 Monday 29 April 2024 (64130) Print this report
Spectral lines during the second week of O4b

I'm repeating the analysis described in #63996 for the second week of O4b. The selected data segment is: 2024-04-20 22:14:40 to 2024-04-21 18:30:12 (72932.0 sec).

As before, I attach the CSV with the identified frequencies and the PDF with the various regions of the spectrum.

Images attached to this report
Non-image files attached to this report
Search Help
×

Warning

×